Specifications
RF7900H high frequency digital x-ray equipment can meet different parts radiography, such as head, chest, abdomen, waist, lumber, thoracic, pelvis, limbs etc. Dynamic Digital Radiology--An enhanced X-ray technology that provides a series of individual digital images acquired at high speed and low dose. It Enables clinicians to observe the dynamic motion of anatomical structures over time, enhancing diagnostic capabilities. Basic functions:Large format (17*17inch) and high-definition radiography, fluoroscopy, real-time video storage and playback, visual contrast, millisecond spot film.
Item | Content | Technical Parameters |
Power | Voltage | 380V±10% |
Frequency | 50Hz±1Hz | |
Capacity | ≥105kVA | |
Internal Resistance | ≤0.17Ω | |
High Frequency High Voltage X-Ray Generator | Power | 50KW |
Inverter Frequency | 500 KHz±20% | |
Radiology Tube voltage | 40kV-150kV ,Steps 1kV | |
Tube current | 10mA-630mA | |
Exposure time | 1ms-10000ms | |
mAs | 0.1-630mAs | |
Fluoroscopy tube voltage | 40kV-125kV | |
Fluoroscopy tube current | 0.5mA-20mA(continues fluoroscopy) | |
5mA-40mA(pulse fluoroscopy) | ||
X-ray source assembly |
Input power | Big focus 74kW Small focus 30kW |
Anode thermal capacity | 220KJ(300KHU) | |
Anode maximum heat dissipation | 750W | |
Rotary anode speed | 8400rpm | |
Tube Focus:big/small | 1.2mm /0.6mm | |
Dynamic WiFi Flat Panel Detector(DRF-1717V) | Active area | 427(H)×427(V) |
Pixel matrix: | 3072(H)×3072(V) | |
Pixel pitch: | 139 μm | |
Limiting resolution | 3.7 lp/mm type | |
A / D transition: | 16 bit | |
Energy range | 40 - 150 kVp | |
Maximum entrance dose (low gain) | 4 mR / frame | |
Power Input: | DC 24V 2A | |
Collimator (XSQ40) | View light: | Light Halogen, AC24V/100W |
Visible light exposure brightness Average illumination brightness: >100Lux | ||
Illuminator turn-on time limit up to 5~45S | ||
Bucky Stand (XPJ20C) | Up and down movement | Not less than 1500mm |
Table (SYC20D) | Lowest height | 550mm |
Table bearing | 200kg |
Registration: routine registration, emergency registration, adding agreement, adding item, clearing information, starting inspection;
Work list: list information, patient search to be examined, refresh to be examined list, delete examination, display column settings. Start inspection and emergency registration;
Check list: list information, checked patient display and search, delete image, image storage, disc burning, add item, display column setting, modify check information;
Patient's body type: thin adult, adult, fat adult;
Photography parameter setting: exposure mode, frame rate setting, kVp, Ma, MS, MAS, AEC, focus selection;
Perspective parameter setting: exposure mode, frame rate setting, kVp, Ma, ABS, time reset;
Browsing tools: zoom, horizontal flip, vertical flip, left turn 90 degrees, right turn 90 degrees, zoom in, zoom out, original size, moving image, reverse color, adaptive size, ROI magnifying glass, magnifying glass, default window width and position, window width and position of interest area, visual window width and position, point gray value, advanced processing, elliptical gray measurement;
System tools: text mark, front position mark, left mark, right mark, circle clipping, delete image, delete tool;
Error reset, exposure indication, full screen, save current image, print
Browsing tools: zoom, horizontal flip, vertical flip, left turn 90 degrees, right turn 90 degrees, zoom in, zoom out, original size, moving image, reverse color, adaptive size, ROI magnifying glass, magnifying glass, default window width and position, window width and position of interest area, visual window width and position, point gray value, advanced processing, elliptical gray measurement, image splicing;
Measurement tools: arrow, cardiothoracic ratio (CTR), distance measurement, angle measurement, spine measurement;
System tools: text mark, front position mark, left mark, right mark, circle clipping, delete image, delete tool;
Report, save current image, print
Report editing: patient information display and editing, photo image selection, report content template selection, report description, report conclusion, report description + conclusion, edit knowledge base, report doctor, audit doctor, report time, print template, setting and saving report;
Report printing: fast printing, printing report
Image archiving, burning, printing: delete image, image storage, browse image, report, lock / unlock, storage queue, print queue;
Disc recording: volume label, save settings, file compression, file structure;
Printing: DICOM printer, local printer
System settings: system, annotation information, tools, others;