Certification: | CE, ISO13485 |
---|---|
Type: | Emergency Treatment |
Display Resolution: | 800 X 600 Pixels |
Sweep Speed: | 12.5mm/S,25mm/S,50mm/S |
Size: | 6.5 Inches (16.5 Cm) Diagonal, Non-Fade |
Lead Selections: | Paddles (Pads), I, II, III, AVR, Avl, Avf, V. |
Customization: |
---|
Suppliers with verified business licenses
Battery
Rechargeable |
12V ;NI-MH |
Capacity |
60 defibrillator discharges @ a new, fully charged battery, or 3 hours minimum continuous ECG monitoring |
Waveform Details:
The tables below provide details of the waveforms delivered by the DM7000 when connected to resistive loads of 25, 50, and 100 Ohms and set to its maximum output. The waveforms are characterized by tyDM7000al values for peak current (Ip), and for monophasic: duration of the output phase (t) or for biphasic: duration of the first output phase (tphase1), and duration of the second output phase (tphase2). The values shown are within 10%.
Waveform: M&B Model DM7000 Truncated Exponential Monophasic
Output Energy Accuracy:
+/- 10% or 1J (whichever is greater) at 50 ohms
+/- 15% or 1J (whichever is greater) at 25 to 100 ohms
Energy Select:
External: 2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 J.
Charge Time:
< 7 seconds @ 360 J (with a SmartPak Plus battery after
15discharges).
<9 seconds @ 360 J (with a SuperPac battery after
15discharges).
< 15 seconds @ 360 J (no battery, 90% AC mains voltage)
Charge Indicator: Audible and graphic.
Output:
Adult paddles, pediatric adapter and multipurpose hands-
free adapter available.
Synchronizer: Delivers energy within 60 msec.
Disarm: Front panel switch.
Monitor/DisplayInput:
3-lead, 5-lead, patient cable, paddles, or multi-
purpose hands-free adapter.
Size: 6.5 inches (16.5 cm) diagonal, non-fade.
Type: color (TFT)
Display Resolution: 800 x 600 pixels.
Sweep Speed: 12.5mm/s,25mm/s,50mm/s
Lead Selections: Paddles (Pads), I, II, III, AVR, AVL, AVF, V.
Frequency Response: (User-selectable).
2 to 20 Hz Limited mode
1 to 40 Hz Monitor mode (before Rev. T8/U8)
0.05 to 150 Hz Diagnostic mode
0.5 to 30 Hz AED mode
Common Mode Rejection:
Meets AAMI EC13-1992, section 3.1.2.1c for 1.2 mV
T-wave (1.0 mV with diagnostic response) and 1mV QRS.